Benjamin, T. A.
Department of Agricultural and Bioresources Engineering, Federal University of Agriculture, Abeokuta
Olufayo, A.
Centre for Research and Development (CERAD), Federal University of Technology, Akure
Omotainse, P. O.
Department of Agricultural and Bioresources Engineering, Federal University of Agriculture, Abeokuta
Dada, P. O.
Department of Agricultural and Bioresources Engineering, Federal University of Agriculture, Abeokuta
ABSTRACT
The calibration of the Delta-T Moisture
Probe was performed using the gravimetric method as a standard. The variation
of soil moisture with depth as determined by the moisture probe and the
gravimetric method was analysed for different plots. Soil samples were obtained
at different depths and the values obtained from the gravimetric method were
compared with those from the Delta-T PR1/6 Moisture Probe. Calibration curves
and general equations with respect to soil at different plots for calibrating
the moisture meter were constructed and a general calibration equation for the
Delta-T PR1/6 Moisture Probe was established. The corresponding generated
values were determined from the established calibration equations. The
corresponding R2 values ranged between 0.8081 – 0.9470. Thus, from
the experiment, moisture probe predictions were high with a range between 80.8
– 94.7% of the variation captured by the gravimetric method. This investigation establishes the fact that the Delta – T
Moisture Probe can be used to carry out soil moisture content determination at
different depths on the field.
Keywords: Soil Moisture content, Moisture Probe, Gravimetric method, Calibration
|
View: 269 | Download: 4
Published
Monday, July 23, 2018
Issue
2018 Special Issue 1
Article Section
GENERAL
The contents of the articles are the sole opinion of the author(s) and not of UJET.
|